Product Details
Custom Matching
Technical Specifications
Measuring item | Value | |
Conductivity | Measuring range | 0.00µS/cm~200.00mS/cm |
Indication error | ±1.0% (F.S) | |
TDS | Measuring range | 0.00mg/L~100g/L |
Indication error | ±0.5% (F.S) | |
Salinity | Measuring range | 0~10000mg/L |
Indication error | ±1.0% (F.S) | |
Temperature | Measuring range | 0 ~99.9℃ |
Indication error | ±1.0 ℃ | |
Testing standard | GB/T 13580.3-1992 | |
Analytical method | Electrochemical method | |
Electrode | Conductivity electrode and temperature electrode | |
Input impedance | ≥1×1012Ω | |
Data storage | Can store 999 set of data records | |
Charging/Data interface | Type C | |
Display | 2.4 inch 240*160 LCD | |
Dimension | 185mm×92mm×50mm | |
Weight | About 0.55kg | |
Working environment | (1) Ambient temperature: 5~40℃ ⑵ Relative humidity: ≤85% ⑶ Power supply: Rechargeable lithium battery or AC (220±22)V (50±0.5) Hz (4) No significant vibration and electromagnetic interference, avoid direct sunlight. |
Functional Features
The measurement process is controlled by a microcontroller unit (MCU) with digital filtering technology to enhance system reliability.
Simultaneous measurement of conductivity and temperature for efficient reading.
Equipped with a high-performance, ultra-low-power microcontroller and a high-capacity rechargeable lithium battery to ensure extended instrument runtime.
Data protection against blackout condition ensures that data is not lost during a power outage.
Storage capacity for 999 data sets with time tags, supporting data protection during power loss and data transfer to a computer for easy record-keeping and analysis.
Auto-lock mode: The display automatically locks onto stable readings.
Equipped with an LCD large screen display for intuitive and convenient operation, supporting one-handed operation and suitable for various usage scenarios.
The instrument is designed with corrosion resistance, waterproofing, and dustproofing to adapt to a variety of harsh environments.
Packing List
Main Unit: 1 unit
Electrode: 1 conductivity electrode, 1 temperature electrode
Reagent: 1 bottle of conductivity calibration solution
5V2A Charger: 1 piece
Carrying case: 1 case
Type C cable: 1 piece
User Manual: 1 copy
Product Certificate: 1 certificate
Warranty Card: 1 card