CHCM-101 CODMn Analyzer

CHCM-101

[Product Overview]

CHCM-101 CODMn analyzer, developed and produced by Shenchanghong, is a rapid, realiable analyzer with high accurate.

[Application Area]

Applicable to the measurement of surface water, tap water, etc.
Product Details
Custom Matching


Technical Specifications

Measuring range

CODMn: 0.00~10.00mg/L

Measuring accuracy

0.01 mg/L

Indication error

≤±5 %

Repeatability

≤±3%

Anti-chlorine interference

≤2000mg/L

Optical stability

Absorbance drift ≤0 .002A/20min; Light source lifespan 100,000 hours

Testing standard

GB/T 11892-89

Analytical method

Closed Vessel Digestion and Colorimetric Method (Colorimetric Tubes)

Processing capacity

20 samples per run

Temperature control system

CODMn digestion temperature: 100℃; Temperature range: 0~99.9℃; Water bath temperature: 100℃;Temperature accuracy:±0.1℃

Water bath period

15min

Curve and Storage

150 factory calibration curves, 20 user-defined curves; Can store 1,000 records.

Display

5-inch LCD screen with a 240*128 pixel matrix

Power

Instrument: 15W

Dimensions

Instrument: 273mm*201mm*129mm; Thermostat water bath: 200mm*200mm*196mm

Weight

Instrument:1.15kg      Thermostat water bath:2.4kg

Working environment

(1) Ambient temperature: 5~40℃ ⑵ Relative humidity: ≤85% ⑶ Power supply: AC 220 (165V-240V) (4) No significant vibration and electromagnetic interference, avoid direct sunlight.

 

 

Functional Features

 

    Utilizes cold light and monochromatic light sources, ensuring exceptional optical stability without interference from various light sources.


    Simple operation with high measurement accuracy, meeting the latest discharge standards.


    Large LCD screen; all settings, calibration, and record operations are performed within the integrated environment.


    Strong anti-interference capabilities using V/F conversion, software redundancy, and software traps.


    Can store working curves and historical records; users can calibrate and store their own curves, with data retention even during power outages.


    Can store 150 factory curves and 1000 records, with data retention even when powered off.


    The main unit casing is made of post-mold ABS material, offering excellent corrosion resistance.

 

Packing List

 

Main Unit: 1 piece

Thermostat water bath: 1 piece

Colorimetric tube: 5 tubes (16*110mm)

Tube rack: 1 piece

Reagent: 1 kit

Power Cord: 2 piece

Manual: 1 copy

Certificate of Conformity: 1 copy

Warranty Card: 1 copy